The new generation of portable photoionization measuring devices.

  • ACI_PPID_01_web.png
  • ACI_PPID_02_web.png
  • ACI_PPID_03_web.png
  • ACI_PPID_04_web.png
  • ACI_PPID_05_web.png
  • ACI_PPID_06_web.png
  • ACI_PPID_07_web.png
  • ACI_PPID_DETAIL_web.png
  • ACI_PPID_SCREEN_web.png
The Portable PID handheld is a new generation of portable photoionization detectors for the detection of volatile organic compounds (VOCs) with an ionization potential < 10.6 eV, which presents the measurement properties known from the PPM in a completely revised, modern concept. The ultra-modern PID technology, combined with the experience of 20 years of measuring device construction, offers the professional user a measuring device with maximum sensitivity and robustness. Designed for field use, the Portable PID is suitable for both personal and stationary measuring tasks where hazardous substance concentrations (gases, vapours) in the air (working environment, indoor, soil gas, environment) need to be monitored. Equipped with a long-life VUV hollow cathode lamp (10.6 eV), the Portable PID covers a large measuring range of 0 - 2,000 ppm (ISOBUTEN equivalent).
The Portable PID meets all the requirements of a modern, low-maintenance and user-friendly photoionization measuring system.

request now

Product Specification

 

Extended measuring range

A new sensor unit allows an extended measuring range of up to 2 000 ppm (isobutene) with high resolution even at low concentrations.

Easy user interface

The high-contrast, graphical 3.5-inch colour TFT display with intuitive user guidance, clear text instructions and step-by-step configuration supports the ease of use even in bad lighting.

High performance PID lamp

The ionization source is a high stable hollow cathode lamp with an energy of 10.6 eV. With its ceramic discharge channel, the stability is improved and a longer service life of up to 15 000 hours and more will achieved.

Acoustic and visual warning functions

The Portable PID has an acoustic alarm and two ultrabright LEDs that warn when adjustable concentration values are exceeded.

Temperature / humidity compensation

The integrated humidity / temperature sensor measures in addition to the sensor signal the physical environment data and compensates their influence on the measuring result.

      

Large data memory

The data memory permits storage of more than 60 000 data records. Additionally, a separate storage is available for manually saved measuring data. A free PC software enables evaluation and logging of the saved data.

Response factors

Response factors for more than 250 gases are stored in the device. Additionally, the user can expand them by up to 20 customer-specific response factors.

Diverse charging options

The Portable PID is equipped with a standard USB connection and can be charged with any common USB wall adapter or car charger.

Easy calibration

The Portable PID is calibrated with 100 ppm Isobuten at delivery. Calibration can be adjusted or repeated when required.

 

 

Technical Data

Detector principle VUV-Photoionization with 10.6 eV hollow cathode lamp
with Ceramic Discharge Channel technology (optional wit 11,8eV lamp)
Detection ranges 0 … 2 000 ppm Isobutene *
Display range 0 … 20 000 ppm, depending on response factor of detected substance
Lower detection limit Typical 10 ppb *
Display resolution Dynamic up to 0.01 ppm
Response time T90 < 10 s *
Signal integrity

Up to 100 ppm typical > 98 % *
Up to 2 000 ppm typical > 95 % *

Influence of humidity Humidity and temperature compensation at 0 ... 50 °C and 0 … 90 % rH
residual effect less than < 10 % FS
Operating conditions 0 … 50 °C 0 … 95 % rH, non-condensing
Storage conditions -20 … 45 °C 0 … 95 % rH, non-condensing
Gas sampling Integrated diaphragm pump (about 300 ml/min), with function monitor
Sample inlet with dust and water protection filter
PID lamp life time 10,6eV: Min. 8 000 hours, typical more than 15 000 h
11,8eV: 4 months from date of delivery
Signalisation Visual: Ultra-bright LEDs (red), Acoustic: Alarm, at least 90 dB (A) in 30 cm
Alarm Locked alarm, Self-triggering alarm, adjustable alarm levels
Display Full graphic color display, 3.5 inch TFT with 320 x 240 pixel
Program functions Continuous measurement with automatic and manual data storage, threshold monitoring, two-point calibration, self-test, fresh air compensation, graphical illustration of the measured data
Memory 60 000 measurement data with time, temperature and humidity
Power supply Integrated lithium ion battery, Charging via USB connection at the PC or wall adapter
Runtime / Charging time Up to 10 hours / < 5 hours with USB charger (100 … 250 V AC / 1 A)
Response factors More than 250 selectable response factors can be set for 10,6eV lamp
Input of user-specific response factors possible for 10,6eV and 11,8eV lamp
Dimension, weight 176,5 mm x 102 mm 39,5 / 49 mm (L x B x H), weight: 550 g
Ingress Protection IP54
Warranty

2 years, except for components in gas path and wearing parts

Approvals EMC directive 2014/30/EU
Scope of delivery Portable PID, USB-charger, USB-cable, USB stick with Windows Software, Carrying case,
User Manual
  * The indicated values were obtained under standardized conditions with 10,6eV lamp. Test gas was isobutene in synthetic air.